free stats Acid Alkylation

Acid Alkylation

Sulfuric Acid Monitoring
Hydrofluoric Acid Monitoring

(click the picture for a high quality version)

Sulfuric Acid Monitoring (top)

A need exists for a reliable, in-situ method to continually monitor sulfuric acid concentration in the alkylation process.  Ideally, a non-contact sensor would be inserted into the acid process in an appropriate location and allowed to communicate with a base instrument located remotely in the safe environment of a control room.  Process Instruments Inc. has started developing a fiber-coupled, Raman scattering instrumentation for this application.  The fiber-optic Raman probe is manufactured out of carbon steel and can be directly inserted into the process stream with no sample conditioning required. Sulfuric Acid Alkylation process
Process Instruments PI-200-L instrument was used to record the Raman spectra of l8 laboratory prepared sulfuric acid dilutions ranging from 95% to 85% acid mixed with water.  The figure to the right illustrates how the Raman spectra change with acid concentration.  The Raman peaks located at 913 cm-1 and 1162 cm-1 both decrease when the acid is diluted while the Raman peak at 1043 cm-1 increases substantially.  The fact that some peaks decrease with decreasing acid concentration and others increase create a very robust chemometric model that is used to correlate the Raman spectra to acid concentrations. Sulfuric Acid
The top spectrum in the figure to the right displays the in-situ spectral response from the spent acid stream that contained a 90 Wt% sulfuric acid solution.  To demonstrate that the spectral features from this on-site measurement were consistent with our freshly prepared samples (discussed above), we also collected the Raman spectrum from a pristine laboratory sample that contained 95 Wt% sulfuric acid.  The bottom spectrum represents a freshly prepared sample.  Similarities in the Raman bands between 850 to 1200 cm-1 demonstrate the ability to obtain real time in-situ Raman spectrum from the spent acid stream that could be used to monitor the acid concentrations within an alkylation unit.
Partial Least Squares analysis of the percent sulfuric acid concentration from 64 alkylation unit samples. The Raman spectrum of each alkylation unit sample along with the laboratory titration measurement were used to make the PLS prediction. The PLS model required three individual factors, and these factors correspond to the three acid peaks in the 850 to 1250 wavenumber region.  The standard deviations of the (predicted minus laboratory) values was 0.11 %Wt, and this variance is within the 0.2%Wt that is inherent in the conventional titration technique.  The observable agreement between the predicted and actual sulfuric acid concentration for the various alkylation unit samples demonstrates that on-line Raman spectroscopy can be utilized for in-situ sulfuric acid monitor.
Hydrofluoric Acid Monitoring (top)

Process Instruments PI-200-L instrument was used to record the Raman spectra of 4 laboratory prepared hydrofluoric acid dilutions ranging from 48% to 19.2% acid mixed with water.  The figure to the right illustrates how the main HF acid peaks located at 655 cm-1 and 759 cm-1 decrease in intensity when the acid is diluted with water.  Hydrofluoric alkylations units require a much higher acid concentration then we could obtain for this study.  Since both of the main HF peaks increase with concentration the elevated alkylation HF acid concentrations will produce an improvement in the signal to noise ratio when compared to this initial study.
The figure to the right provides an expanded view of the frequency shift axis to also display the observed water Raman band located around 1630 cm-1 that increases in intensity when the acid is diluted with water.