Raman Instrumentaion
Diagram of the System
Sample Characteristics
Instrumentation Stability
Instrument Calibration
(Click on pictures for a high quality version)
|
Diagram of the System Top
|
 |
|
|
A diagram of the components of the fiber-coupled diode laser Raman system. The components include: the computer and interface electronics, the diode laser, the f/1.4 spectrograph, the CCD array, the collection fiber bundle, and the remote sample cell.
|
|
|
|
|
|
|
Sample Characteristics Top
|
|
|
|
|
| Spectrum Resolution High vs. Low in Gasoline |
|
|
|
 |
|
| A spectrum of unleaded premium gasoline obtained with 2 wavenumber resolution compared with a gasoline spectrum obtained with 30 wavenumber resolution. The individual components easily distinquishable with 2 wavenumber resolution are washed out at 30 wavenumber resolution |
|
Near-IR Absorbance Spectra: Five Unleaded Gasolines |
|
|
|
|
|
 |
|
| Five near-IR absorbance spectra of different unleaded gasolines. The change in the spectra are very small requiring chemometrics to pull the differences out of the data. |
|
Unleaded Premium Gasoline Spectrum: High Resolution Spectrum |
|
|
|
|
|
 |
|
| A spectrum of unleaded premium gasoline showing some primary components. The primary components shown are: m-xylene, o-xylene, iso-octane, ethyl benzene, toluene, p-xylene, and benzene. |
|
Sample Fluorescence: Unleaded Premium Gasoline |
|
|
|
|
 |
|
| Comparison of a spectrum of unleaded premium gasoline for a variety of wavelengths from 488 nm to 852 nm. The fluorescence background of the gasoline sample decreases significantly with increasing excitation wavelength. |
|
|
|
|
|
|
|
|
|
|
Instrumentation Stability Top
|
System Stability, 4 Weeks, 7500 Spectra |
|
|
|
50/50 Toluene / Acetonitrile Solution and a Diamond Reference Standard |
|
 |
|
|
Peaks Fit to a Gaussian / Lorentzian Combination |
|
|
|
 |
|
|
System Stability, 4 Weeks, 7500 Spectra |
|
|
|
 |
|
|
|
Histogram of Fitted Gaussian / Lorentzian Mean Toluene : 1003.445 +/- 0.011 |
|
|
|
 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Instrument Calibration Top
| PI-200 Neon Calibration Lamp with Andor CCD Camera |
|
|
|
 |
|
|
| Spectrum of a Neon Calibration Lamp taken with the PI-200 Raman analyzer and the Andor CCD camera. We use this spectrum to calibrate the pixels of the CCD Camera in wavelength (nm). |
|
| PI-200: Toluene Acetonitrile Calibration |
|
|
|
 |
|
| Spectrum of Toluene Acetonitrile taken with the PI-200 Raman analyzer and the Andor CCD camera. We use this spectrum to calibrate the pixels of the CCD Camera in wavelength (nm). |
|
| Intensity Calibration with Broadband Source |
|
|
|
|
 |
|
| Spectrum of a Broadband Source. We use this spectrum to calibrate the intensity of the CCD Camera in the Raman Instrument. |
|
|
|
|
|
|
|
|
|
|
|